Table 3 Duration of wear (first to last valid wearable vital signs measurement), completeness of data capture and reasons for wearable removal. The blood pressure cuff was not applied to 15 participants at their request

From: The completeness, accuracy and impact on alerts, of wearable vital signs monitoring in hospitalised patients

 

Wearable Sensor (N = 48)

HR/RR (LifeTouch)

Temperature (LifeTemp)

SpO2 (Nonin PulseOx)

BP (A&D TM2441)

Duration of sensor wear (days/participant)

3.9 [1.7, 5.9]

3.9 [1.6, 5.9]

3.8 [0.9, 5.9]

1.9 [0.9—3.8]

Overall completeness (%) of wearable sensor data

81.2

92.1

68.6

38.4

Completeness (%) per participant

83.8 [64.1, 95.7]

97.7 [79.7, 99.8]

72.3 [61.7, 87.2]

35.8 [16.3, 47.6]

Reason for device removal

 Discharge from hospital

25 (52.1)

23 (47.9)

24 (50.0)

9 (18.8)

 Sedated/ventilated on ICU

4 (8.3)

4 (8.3)

3 (6.3)

-

 Participant request

19 (39.6)

21 (43.8)

21 (43.8)

22 (45.8)

 Never applied

-

-

-

15 (31.3)

 Other

-

-

-

2 (4.2)

  1. BP blood pressure, SpO2 oxygen saturation